Observation Possible Cause Action
A bubble was present in the
Sample Loading Blade when it
was used to apply the PCR
reaction to the Digital PCR 20K
Chip.
If possible, use the
AnalysisSuite
™
Software to filter
the low quality data points, or
discard the chip and run the
sample again.
When loading the Sample
Loading Blades:
• If you are using a manual
pipette, pipette to the first
stop.
• Decrease your pipetting
speed.