Observation Possible Cause Action
A bubble was present in the
Sample Loading Blade when it
was used to apply the PCR
reaction to the Digital PCR 20K
Chip.
If possible, use the
AnalysisSuite
Software to filter
the low quality data points, or
discard the chip and run the
sample again.
When loading the Sample
Loading Blades:
If you are using a manual
pipette, pipette to the first
stop.
Decrease your pipetting
speed.
Chapter 5 Troubleshooting
Troubleshooting chip images using the Chip View
5
80
QuantStudio
®
3D Digital PCR System User Guide