Analyze the chips
The following materials are required to image loaded Digital PCR 20K Chips on the
QuantStudio
®
3D Digital PCR System.
• Prepared Digital PCR 20K Chip (from the previous section)
• Isopropanol
• Gloves, powder-free, nitrile
• Cleanroom-grade, low-lint polyester wipes
For optimal efficiency when running large numbers (>6) of Digital PCR 20K Chips,
follow the guidelines below:
• Run the Digital PCR 20K Chips in alphanumeric order (according to chip ID) to
help avoid mistakes when entering sample information.
• If you are running multiple chips, you can run the next chip in the series when
the touchscreen displays the Analyzing Chip screen .
To prepare the instrument for imaging data, edit data destination and file name
settings before loading a chip into the instrument.
1.
Set the destination for the imaging data:
a.
From the Main Menu of the QuantStudio
®
3D Instrument, touch , then
touch Data Destinations.
b.
In the Data Destinations screen, touch the desired destination for the
imaging data.
You can configure the QuantStudio
®
3D Instrument to upload the imaging
data to a USB drive, a network file server, or the cloud data storage service.
Materials
required
Guidelines for
high-throughput
analysis
Prepare the
instrument for
imaging data
Chapter 4 Analyze the Prepared Chips
Analyze the chips
4
QuantStudio
®
3D Digital PCR System User Guide
69